Phi96+384.fr
FLuorometric Imaging Plate Reader
Measuring Intracellular pH
With the FLIPR® I and FLIPR384
Fluorometric Imaging
Plate Reader Systems
Section I: Introduction
The FLIPR® system is designed to perform functional cell-based assays. This application note provides a basic protocol for running an intracellular pH assay on the FLIPR system using adherent cells, as well as a discussion of some of the important parameters for optimization of the assay. Because each cell line has unique properties, the protocol will need to be optimized for your particular assay.
The "Principles" section of this application note outlines the purpose and the options for each step of the assay. The next section details the materials and methods used in the assay.
Section II: Principles of the intracellular pH assay
The Na+/H+ exchanger, or NHE, contributes to the pH stability of the cytoplasm by exporting protons out of the cells and importing sodium ions. BCECF is a pH sensitive dye. The extinction coefficient of BCECF is high in alkaline environment and low in acidic environment. It can be used to monitor the NHE activity by measuring changes in intracellular pH following an artificial acid-load of the cells. Cells are initially incubated in the presence of BCECF, then NH4Cl, which
causes the dye to enter the cell and the cytoplasm to alkalinize due to the formation of NH3 (high fluorescence signal). When NH4Cl is diluted by the
addition of fluid on the FLIPR system, NH3 diffuses out of the cell, leaving high
concentration of H+ inside the cytoplasm (resulting in rapid decrease of fluorescence signal). These protons are removed from the cells by the action of the NHE (demonstrated by a slow increase of fluorescence signal, proportional to the NHE activity). Figure 1 represents typical results obtained with the intracellular pH assay.
The activation of many G-protein coupled receptors causes acid excretion, mainly through the NHE activity. It is possible to measure the response of such receptors to ligands by monitoring the activation of the NHE in cells subjected to the BCECF and acid load treatment. The intracellular pH assay may provide an alternative to intracellular calcium assay for non-Gq-protein coupled receptors. This assay may also be used to assess inhibitors of the NHE function.
Cells are alkaline
0.01 µM carbachol
Cellular recovery from
acidificationdue to NH3
intracellular acidificationactivity of NHE
leaving the cells
Figure 1: Drawing of intracellular pH results using M1-WT3 cells and carbachol agonist.
Note: This application note provides a protocol for adherent cells only, since the volume added on the FLIPR system requires the cells to be firmly attached to the bottom of the plate.
Cells are seeded the day before the experiment and all steps are carried out in the same black wall 96- or 384-well plate. The basic steps of the assay are:
•Plate the cells in black-wall 96- or 384-well plate.
•Dye load the cells with BCECF for 45 minutes.
•Prepare the acid load of the cells by adding 20 mM NH4Cl into the dye load-
ing plate, incubate for 15 minutes.
•Wash the cells of extracellular dye with wash buffer containing 20 mM
•Assay the cells in the FLIPR system; the NH4Cl buffer is diluted by the addi-
tion of a large volume of balanced salt solution and various ligands/controls.
It is necessary to optimize the cell seeding density so that a uniform, confluent monolayer is formed after an overnight incubation. Cells can be seeded more than one day in advance, as long as the seeding densities are adjusted to yield a confluent monolayer on the day of the assay. Cells that are weakly adherent or do not grow uniformly may need to be grown on plates coated with a matrix such as poly D-lysine, laminin or collagen (see Appendix for details). Cells that are normally maintained in culture at subconfluent levels should be seeded at lower densities.
DYE LOADING
In order to observe changes in intracellular pH, cells must be "loaded" with a pH-sensitive fluorescent dye. The dye loading protocol must be optimized for each cell type.
Fluorescent dye
To date, the most commonly used dye for intracellular pH assay has been BCECF.
Anion exchange protein inhibitor
Some cell types use mechanisms such as the anion exchange protein to export
anionic molecules from the cells, including anionic forms of the fluorescent dyes.
Not only will this result in poor dye loading, but it also may cause an artifact in
the data; there will be a sharp decline in the measured fluorescence when the test
compounds are added, due to dilution of the extracellular dye. (The test com-
pounds are prepared in buffer without dye.) Therefore, it may be critical to the
success of the FLIPR system intracellular pH assay to inhibit the action of the
anion exchange protein.
Probenecid is an anion exchange protein inhibitor and when added to the loading medium may increase dye retention in the cells. An example of a cell type known to require probenecid is CHO. Although probenecid can be useful in slowing dye leakage from cells, it is toxic to the cells, and hence the duration of the dye loading should be kept to a minimum (see Loading duration).
Sulfinpyrazone is another anion exchange inhibitor. To date, little information about using sulfinpyrazone in the intracellular pH assay is available.
Dye loading media
Several types of loading media have been tested. The medium should be opti-
mized for both satisfactory dye loading and good cellular responses. Possible
choices include:
•Growth medium + 10% fetal bovine serum (FBS) + 20 mM HEPES•Growth medium + 1% FBS + 20 mM HEPES•Hank's BSS 1X (without phenol red) + 20 mM HEPES and 1% FBS•Hank's BSS 1X (without phenol red) + 20 mM HEPES and 1% BSA
Loading duration and temperature
The optimal loading time will depend on the cell type. Because the fluorescent
dye is toxic to cells, it is best not to exceed the optimal loading time. A 60-minute
loading time at 37°C is usually effective for most cells and is the recommended
starting point for assay development.
Note: If loading for 30 minutes yields an acceptable fluorescence signal, use the shorter loading time.
PREPARING FOR
This step is actually not the acid load itself, but the addition of NH4Cl to the cells.
THE ACID LOAD
Once in the cells, NH4Cl forms NH3 and alkalinizes the cytoplasm. Subsequently,
OF THE CELLS
when NH4Cl is diluted out by fluid addition on the FLIPR system, NH3 diffuses
out of the cells, leaving the protons H+ in high concentration in the cytoplasm. This triggers the activity of the NHE and the export of protons out of the cell.
The solution made for this step is a 200 mM NH4Cl solution. NH4Cl remains in
the wells throughout the wash step until diluted on the FLIPR system.
PREPARING THE
Depending on its complexity, preparing a compound plate can take a variable
COMPOUND
amount of time. Therefore, it is best to plan your experiment carefully to ensure
that the compound plate will be ready for use as soon as the cells are ready. In this application note, it is assumed that the dye loading incubation is long enough to prepare the compound plate.
Preparing the compound diluent
In the intracellular pH assay, the buffer used to dilute the compounds and the
wash buffer are different (in contrast to intracellular calcium and membrane
potential assays). The wash buffer contains 20 mM NH4Cl, whereas the com-
pound diluent contains no NH4Cl.
You can prepare a large volume of compound diluent and use it for all of the plates you will be assaying in a day. Using V-bottom plates to hold the compounds will minimize the dead volumes in the compound plate.
Preparing the compounds
The compounds should be prepared at 1.25X the final concentration in the cell
plate during the assay. Adding a large volume of compound is essential for this
assay since it allows the acidification of the cytoplasm, triggering the activation of
the NHE.
Intracellular pH measurements are performed quickly in order to visualize rapid cellular kinetics. Therefore, it is necessary to pipet the compounds into the cell plate rapidly to effectively mix the fluids in the wells. The large volume added and the fast pipetting speed provide instantaneous mixing.
Because of the volume of compound added to the cell plate, strongly adherent cells are better candidates for this assay than weakly adherent ones.
WASHING THE
After dye loading and addition of NH4Cl solution to the plate, several washes
with the wash buffer are necessary to remove extracellular dye. The washes are performed in the presence of 20 mM NH4Cl. An adequate automated cell washer
should have adjustable dispensing and aspirating heights and speeds. We recommend Molecular Devices cell washer for 96- or 384-well plates. Alternative washers are BioTek or LabSytems cell washers for 96-well plates, and Skatron Embla340 for 384-well plates.
The quality of data obtained from the FLIPR system will depend partially on the cell washer's ability to leave a consistent residual volume in each well after the last wash. Results obtained after manual washing tend to be more variable. Also, if washing the cells manually, it is critical not to aspirate the cells dry between washes.
Note: The wash buffer is NOT the same as the buffer used to dilute the compounds.
FLIPR SETUP
Checking the background and basal fluorescence signals
PARAMETERS FOR
Start with the laser power, camera F-stop and exposure time set as shown in
PH ASSAYS
0.300 W - 0.400 W
0.600 W - 0.800 W
Table 1: FLIPR hardware settings for a cell plate basal fluorescence signal test.
At the beginning of a data run, perform a "signal test" to check the background fluorescence and the basal fluorescence signal (i.e., prior to compound addition) from the cells. For intracellular pH assays, it is best to work with a basal fluorescence signal of 40,000 to 50,000 counts above background (saturation of the camera occurs at 65,000 counts). The difference in laser power used on the FLIPR I system and on the FLIPR384 system is due to the difference in the way the excitation light scans the plate. The laser power used on the FLIPR384 sytem is typically twice as high as on FLIPR I to obtain the same basal fluorescence signal.
Basal fluorescence signal The desirable basal fluorescence signal needs to be high enough to allow the drop of fluorescence after the fluid addition on the FLIPR system. However, it should not be so high that the camera reaches saturation.
Adjusting the basal fluorescence signal
Adjust the following instrument parameters to obtain an acceptable basal fluores-
cence signal:
Exposure timeIf the basal fluorescence signal is too high, the exposure time can be decreased to a minimum of 0.1 second. If the basal fluorescence signal is too low the exposure time can be increased, but the sampling interval will have to be increased to at least the exposure time + 0.6 second (it takes 0.6 seconds for the camera to inte-grate data).
Camera F-stopIncreasing the camera F-stop number decreases the opening of the aperture. If the basal fluorescence signal is too high, increase the F-stop. If the basal fluorescence signal is too low, the laser power may be increased. F/1.4 is the largest aperture opening setting for the FLIPR system.
Laser powerIncrease the laser power if the basal fluorescence signal is too low, or decrease it if the basal fluorescence signal is too high. The laser power during the assay should range between 0.1500 W and 0.800 W on FLIPR I, 0.300 W and 2.000 W on the FLIPR384 system.
PIPETTOR
The FLIPR system software allows you to set the height of the pipet tips in the
PARAMETERS FOR
wells during fluid transfer, as well as the speed at which fluid is dispensed. These
parameters are set independently for each pipetting sequence.
PH ASSAYS
Pipettor height
After the pipettor picks up fluid from the compound plate, it draws a small bub-
ble of air into the bottom of each pipet tip to ensure that fluid doesn't leak out.
The bubble will be the first thing out of the pipet tip when fluid is dispensed. To
avoid blowing bubbles in the wells (which can cause random light reflections and
spurious signals), it is best to start dispensing with the tips above the fluid level in
the wells. It is also preferable to have the pipettor tips submerged after the addi-
tion has been completed to ensure that all the sample is dispensed. If the pipet
tips are in the air at the end of the fluid delivery, a drop can form on the end of the
tip due to surface tension. Therefore, the pipettor height before fluid addition
should be somewhere above the starting level of fluid in the wells but below the
final volume after the addition. For example, if the wells contain 100 µL and the
sample volume to be added is 50 µL, the pipettor can dispense the compounds
from a height of 120–140 µL.
Fluid dispensing speed
The default pipettor dispense speed is 50 µL/sec for 96-well plates and 20 µL/sec
for 384-well plates. Table 2 provides the limits for the fluid dispense speed for
both 96- and 384-well plate formats.
Slow dispense speed
10 µL/sec - 40 µL/sec
5µL/sec - 10 µL/sec
(weakly adherent cells)
Fast dispense speed
50 µL/sec - 80 µL/sec
15 µL/sec - 20 µL/sec
(strongly adherent cells)
Table 2: Fluid dispense speeds.
These values must be determined experimentally for each cell type, but generally it is preferable to dispense as fast as possible to enhance mixing of the compounds in the wells. The trade-off is that the pipetting speed must not be so forceful that it dislodges cells from the well.
Section III: Method for adherent cells
PREPARING THE
Adherent cells are typically plated the day prior to the experiment.
Clear, flat-bottom, black-wall 384- or 96-well plates (see consumables list in Appendix B). The flat bottom ensures that the cellular fluorescence is localized to a single horizontal plane, while the black walls prevent well-to-well crosstalk.
Cells ready to be transferred into 96- or 384-well plates (see Table 3).
Incubator (5% CO2, 37 °C).
Pipettor and sterile tips suitable for use with microplates.
Plating the cells
2x104 - 1x105 cells/well
1x104 - 5x104 cells/well
Total cell number
2x106 - 1x107 cells
4x106 - 2x107 cells
Growth medium volume
30 - 80 µL/well,
(12 - 32 mL/plate)
Table 3: Recommended cell seeding densities for adherent cells.
Prepare a cell suspension in growth medium (see cell densities in Table 3).
Pipette the cell suspension in plate (see volume per well in Table 3).
Incubate the cell plate overnight in a 5% CO2, 37°C incubator.
Note: To minimize edge effects, avoid stacking cell plates in the incubator.
DYE LOADING
Preparing the dye
The 1 mg/mL stock of BCECF dye is mixed with an equal volume of 20% (w/v) pluronic acid immediately before use.
Note: Some cell lines generate better results when pluronic acid is not used in the loading
medium.
1 mg/mL (= 1.6 mM) BCECF stock. (See Appendix B for consumables details.)
Solubilize 1 mg BCECF in 1 mL low-water DMSO.
Store 25 µL aliquots at -20°C.
20% pluronic acid solution:
Note: A ready-to-use 20% pluronic acid solution is available from Molecular Probes.
Weigh out 400 mg pluronic acid into a tube, then solubilize in 2 mL low-water DMSO. Warming the solution to 37 °C will increase the acid solubility. Mix the solution gently to avoid forming excess bubbles.
Allow the solution to cool to room temperature before aliquoting or using, and store 25 µL aliquots at room temperature.
Dye/pluronic acid mixture:
Immediately before use, combine equal volumes of the dye stock and 20% pluronic acid. The dye and pluronic acid will be at concentrations of 0.8 mM and 10%, respectively.
Preparing the anion exchange protein inhibitor
Note: probenecid is only required for some cell types, e.g., CHO.
Probenecid should be prepared fresh every day at a stock concentration of 250 mM, and used at a working concentration of 2.5 mM.
250 mM probenecid (100X stock, enough for 1L of buffer):
Solubilize 710 mg probenecid in 5 mL 1.0 N NaOH.
Mix in 5 mL Hank's BSS 1X without phenol red with 20 mM HEPES.
Preparing the loading medium
Possible choices of loading medium include:
•Growth medium + 20 mM HEPES + 10% fetal bovine serum (FBS)•Growth medium + 20 mM HEPES + 1% FBS •Hank's BSS 1X (without phenol red) + 20 mM HEPES and 1% FBS•Hank's BSS 1X (without phenol red) + 20 mM HEPES and 1% BSA
Note: HEPES is typically added to Hank's buffer and /or any loading medium where probenecid is present in order to maintain constant pH.
Dye loading the cells
Using a cell washer instrument (for 96- or 384-well plates)
BCECF/pluronic acid mixture
88 µL (6-10 µM dye final
104 µL (6-10 µM dye final
(2X for cell washer method)
Table 4: Preparation of loading medium for one plate of adherent cells, using the cell washer.
Wash off the growth medium using a cell washer. For most cell lines, one to two washes is sufficient. Make sure the residual volume after the last wash is defined and consistent across the plate. The recommended
residual volumes are 45 µL for 96-well plate and 17.5 µL for 384-well plate.
Add an equal volume of a 2X concentration of loading buffer using a multi-channel pipettor. The recommended final volumes in the wells are 90 µL for 96-well plate and 35 µL for 384-well plate.
Cover the plate, then incubate in a 5% CO2, 37°C incubator for 45
The cell washer method is effective at not perturbing the cell layer. It is strongly recommended to use the cell washer method only for 384-well plates, since the entire well area is read by the FLIPR384 system. For 96-well plates, you can use either a cell washer or the manual method.
Using the manual method (for 96-well plates only):
BCECF/pluronic acid mixture
44 µL (3-5 µM dye final
52 µL (3-5 µM dye final
(1X for manual wash method)
Table 5: Preparation of loading medium for one plate of adherent cells, using the manual method.
Aspirate all of the growth medium out of the wells, being careful to avoid disturbing the cells in the area read by the FLIPR system.
Area read in 96-well plate
Figure 2: The region in a microplate well read for data in a 96-well plate.
Dispense the loading medium into each well. The recommended volumes are 90 µL for a 96-well plate and 35 µL for a 384-well plate.
Cover the plate, then incubate in a 5% CO2, 37°C incubator for 45
PREPARING FOR
The acid load solution is a 200 mM NH4Cl solution (10X concentration).
THE ACID LOAD
OF THE CELLS
Preparing the acid load solution (buffer/200 mM NH4Cl)
Prepare a stock in H2O: 535 mg NH4Cl + 10 mL H2O = 1 M stock
Dilute this stock in buffer to make the 200 mM NH4Cl acid load
40 mL Hank's BSS 1X
800 µL HEPES (1 M)
10 mL (1 M) NH4Cl stock. New stock concentration is 200 mM.
Optional, if required by your cells:
400 µL (250 mM) probenecid.
Preparing for the acid loading of the cells
Volume during dye loading
NH4Cl 200 mM added
Table 6: Recommended volumes during dye loading and acid load.
Approximately 45 minutes after the beginning of the dye loading incu-bation, add the 200 mM NH4Cl solution to all wells: 10 µL for a 96-well
plate and 5 µL for a 384-well plate. The final concentration of NH4Cl is
Transfer the plate to the 37°C incubator for an additional 15-20 minutes.
PREPARE THE
Preparing the compound diluent
COMPOUND
The compound plate should be prepared during the dye loading incubation. The
compound diluent
does not contain NH4Cl.
For 100 mL compound diluent:
100 mL Hank's BSS 1X (without phenol red).
2 mL (1 M) HEPES.
Optional, if required by your cells:
1 mL (250 mM) probenecid.
Note: If compounds are proteins or peptides, it may be necessary to add 0.1%BSA to the wash and the compound diluent buffer to prevent the compounds from sticking to the plastic of the plate.
Preparing the compounds
The compound concentration is determined by the necessity to dilute out the
NH4Cl in the wells. Therefore, compounds are prepared at a 1.25X concentration.
WASHING THE
Preparing the wash buffer (buffer/20 mM NH4Cl)
For 500 mL wash buffer containing 20 mM NH4Cl (1X concentration), enough for two to three 96- or 384-well plates:
500 mL Hank's BSS 1X (without phenol red).
10 mL (1 M) HEPES
50 mL NH4Cl 200 mM (final concentration 20 mM)
Optional, if required by your cells:
5 mL (250 mM) probenecid.
Washing the cells
You will need approximately 150 mL/96-well plate or 200 mL/384-well plate of the wash buffer. Dead volumes for cell washers vary and must be determined by the user for their cell washer.
Wash the cells three to four times with the cell washer. Use a "gentle" setting if the cells are weakly adherent. Leave a defined, consistent volume of fluid covering the cells after the last wash. For the intracellular pH assay, the residual volume should be as small as possible to allow a large volume to be added on the FLIPR system to dilute out NH4Cl. The recommended volume ratio is 4:1 (4 parts of
volume added on FLIPR to 1 part of volume in cell plate). Some washers do not permit such a small volume; you may also leave a large volume after the wash, and manually pipette out of the wells a defined volume to reduce the final volume after the wash.
Final volume after wash
Volume added on the FLIPR
Table 7: Recommended volumes in assay.
RUNNING THE
Start with the laser power set at 0.300 Watts for the FLIPR I system, 0.600 Watts for the FLIPR384 system, the exposure time set to 0.4 seconds, and a camera F-stop of F/4 or 5.6. At the beginning of a data run, perform a "signal test" to check the basal fluorescence signal from the dye-loaded cells. If it is not between 40,000 to 50,000 counts above background, adjust it by modifying the exposure time, the camera F-stop, and the laser power.
EXAMPLE OF
The following FLIPR system setup parameters are for an assay using CHO cells
PH ASSAY SETUP
PARAMETERS
Dialog Box
Settings for
Settings for 384-
well Plates
General Tab
Residual Volume in Cell Plate
Compound Concentration
Multiple Sequences
Neither activated
Neither activated
Sample Interval = 1.0 sec
Sample Interval = 1.0 sec
Sample Count = 60
Sample Count = 60
Sample Interval = 3.0 sec
Sample Interval = 3.0 sec
Sample Count = 20
Sample Count = 20
Fluid Addition Tab
Mix After Addition
Number of Mix Cycles
Remove Fluid After Addition
Appendix A: Improving confluence and/or adherence
of weakly adherent cells
If you are working with a cell line that doesn't grow uniformly on microplates, or that adheres weakly, coating the bottom of the cell plate with a matrix (i.e, poly D-lysine, laminin, fibronectin, gelatin, etc.) may improve the homogeneity of the monolayer and may help weakly adherent cells to remain attached to the plate through the washing process. Sample procedures for poly D-lysine, laminin, and collagen follow.
a) Poly D-lysine Coating (or Becton Dickinson part # 354640)
Prepare a sterile, 100 µg/mL solution of poly D-lysine (Sigma Chemical Co., catalog # P7280) in tissue culture grade water. You will need approximately 8 mL per 384-well plate or 5 mL per 96-well plate to be coated.
Working in a tissue culture hood, aliquot approximately 20 µL into a 384-well plate or 50 µL into a 96-well plate, then leave the plate(s) in the hood for 30 minutes.
Aspirate off the poly D-lysine with a Pasteur pipet, then rinse the plate once with sterile water: 100 µL/well for a 384-well plate or 200 µL /well for a 96-well plate. Aspirate off the water.
Allow the plates to dry in the hood before use.
b) Laminin Coating
Prepare a sterile, 66.7 µg/mL (500 µg/7.5 mL) solution of laminin (Sigma Chemical Co., catalog # L-6274) in Hank's BSS. You will need approximately 8 mL per 384-well plate or 5 mL per 96-well plate to be coated.
Working in a tissue culture hood, aliquot approximately 20 µL/well for a 384-well plate or 50 µL /well for a 96-well plate, then leave the plate(s) in the hood for 30 minutes.
Aspirate off the laminin with a Pasteur pipet.
Allow the plates to dry in the hood before use.
c) Collagen Coating (or Becton Dickinson part # 354649)
Prepare a sterile, 0.3 mg/mL solution of Collagen (Vitrogen 100, from Collagen Biomedical, Palo Alto, California) in sterile 0.01 N HCl. You will need approximately 8 mL per 384-well plat or 5 mL per 96-well plate to be coated.
Working in a tissue culture hood, aliquot approximately 20 µL /well for a 384-well plate or 50 µL for a 96-well plate, then leave the plate(s) in the hood overnight.
Aspirate off the collagen with a Pasteur pipet.
The next day, rinse once with sterile PBS 1X to neutralize the pH.
Appendix B: List of consumables
Item number
Black wall plates, clear bottom, tissue
culture treated, sterile, 96-well
Packard Instrument
Black wall plates, clear bottom, tissue
culture treated, sterile, 384-well
Greiner/distributor
Nunc V-bottom 96 well plate
Nunc part #249128
Non-sterile lids for Nunc plates
Clear plate, 384 well, for compounds,
Greiner/distributor
FLIPR pipet tips, black, non-sterile
Molecular Devices
(for experiments), 96-well
FLIPR pipet tips, black, non-sterile
Molecular Devices
(for experiments), 384 well
Aspirator manifold (12 pin)
Aspirator manifold (8 pin)
Hank's Balanced Salt Solution
HEPES buffer solution 1 M
Irvine Scientific
Probenecid, crystalline
Pluronic acid 20% solution
BCECF AM ester for intracellular pH
DMSO, low water content
TRADEMARKS: FLIPR is a registered trademark of
Molecular Devices Corporation
Source: http://labs.pbrc.edu/cellbiology/documents/pHBCECF.pdf
Journal of Cell and Molecular Research (2013) 5 (1), 24-34 Molecular docking approach of monoamine oxidase B inhibitors for identifying new potential drugs: Insights into drug-protein interaction discovery Department of Biology, Faculty of Science, Shahrekord University, Shahrekord, Iran Received 16 August 2013 Accepted 14 September 2013
Under the Patronage of H. H. Sheikh Nahayan Mabarak Al Nahayan Minister of Higher Education and Scientific Research, Chancellor of the UAE University The College of Food and Agriculture, UAE University and Municipalities and Agriculture Department, Agricultural Sector First conference of the International Society of Camelids